Overview
  • Intended Use

    Safe Accu2 is intended for IVD testing the blood glucose in the fresh whole capillary blood and vein plasma; for medical institutes’ fast testing of blood glucose; for diabetics and other groups’ blood glucose monitoring.

  • Test Principle

    C6H12O6 + K3Fe(CN)6   →   FAD-GDH   →  C6H12O7 + K4Fe(CN)6
                        K4Fe(CN)6 → K3Fe(CN)6+ e-

  • Features (Technical advantages)

    1) FAD-GDH system : It is not easily affected by oxygen molecule in the blood.

    2) User-friendly operation system: no coding & automatic test strip ejectionin the management of diabetes.

Specification
Blood volume0.6μL
Sample typeCapillary whole bloodyenous whole blood
CalibrationPlasma equivalent
Measuring time10s
Meter storage/transportation condition-20C~55℃
Dimension105*60*23(mm)
Weight77.2g,with battery
Power source3V DC,10mA,2 AAA alkaline batteries
Memory200 blood glucose test results
Operating condition10℃~35C≤80% RH
ConstructionHand-held
Measurement unitsmg/dL or mmol/L
Measurement range20~600 mg/dL or 1.1~33.3mmo/L
Precision: test result of Safe-Accu2 meet the requirement below:

Accuracy: 95% test result of Safe-Accu 2 meet the requirement below:


Concentration RangeBias%
<5.5 mmol/L(100 mg/dL)Within±0.83 mmol/L(15 mg/dL)
25.5 mmol/L(100mg/dL)Within ± 15%


Precision: test result of Safe-Accu 2 meet the requirement below:


Concentration RangeRequire ment
<5.5 mmol/L(100 mg/dL)SD<0.34mmol/L(6.0mg/dL)
25.5 mmol/L(100mg/dL)CV<6.0%



                         