Safe AQ Smart
Fast 5 secondsTest Time
Sinodraw Patent Lancing Device

Overview
Overview
The Safe AQ Smart Blood Glucose Monitoring System is designed for the quantitative measurement of glucose in fresh capillary whole blood samples taken from the fingertip and in venous whole blood samples. The Blood Glucose Monitoring System is for use outside the body only (in vitro diagnostic use) for self-testing and professional use as an aid in the management of diabetes.
Test principle
A blood glucose test is based on measurement of electrical current caused by the reaction of glucose with the reagents (special chemicals) on the electrode of the strip. The blood or control solution sample is drawn into the tip of the test strip through capillary action. Glucose in the sample reacts with the special chemicals and generates electrons, which produce electrical current. The blood glucose meter measures the electrical current and calculates the glucose result. The glucose result is displayed by your meter as mg/dL or mmol/L.
Features (Technical advantages)
1) FAD-GDH system : It is not easily affected by oxygen molecule in the blood.
2) User-friendly operation system: no coding & automatic test strip ejectionin the management of diabetes.
Specification
Blood volume | 0.6μL |
Sample type | Capillary whole bloodyenous whole blood |
Calibration | Plasma equivalent |
Measuring time | 5s |
Meter storage/transportation condition | -20℃~55℃ |
Dimension | 84*60*26(mm) |
Weight | 73.5g, with battery |
Power source | 3V DC, 50mA, 2 AAA alkaline batteries |
Memory | 10 blood glucose test results |
Operating condition | 10℃~35℃; ≤80% RH |
Construction | Hand-held |
Measurement units | mg/dL or mmol/L |
Measurement range | 20~600 mg/dL or 1.1~33.3mmo/L |
Shelf life | 10 years(estimated by test 7 times per day). During use, user should maintain the product refer to this user manual’s requirements. |
Performance Characteristics
Accuracy: 95% test result of Safe AQ meet the requirement below:
Concentration Range | Bias% |
---|---|
<5.5 mmol/L(100 mg/dL) | Within±0.83 mmol/L(15 mg/dL) |
25.5 mmol/L(100mg/dL) | Within ± 15% |
Precision: test result of Safe AQ Smart meet the requirement below:
Concentration Range | Require ment |
---|---|
<5.5 mmol/L(100 mg/dL) | SD<0.34mmol/L(6.0mg/dL) |
25.5 mmol/L(100mg/dL) | CV<6.0% |