Safe AQ Voice
Fast 5s test time & Patent painless lancing device
Non-interference & Voice reminder & Red-green light warning.

Overview
Safe AQ Voice Blood Glucose Monitoring System is intended for in vitro diagnostic use and should not be used for the diagnosis of or screening of diabetes.
Package
The Safe AQ Voice Blood Glucose Monitoring System includes:
Safe AQ Voice blood glucose meter, Safe AQ blood glucose test strip, blood glucose control solutionFeatures (Technical advantages)
1) FAD-GDH system: It is not easily affected by oxygen molecule in the blood.
2) User-friendly operation system: fast 5s test time & patent painless lancing device & non-interference & voice reminder & red-green light warning.Test principle
C6H12O6 + K3Fe(CN)6 → FAD-GDH → C6H12O7 + K4Fe(CN)6
K4Fe(CN)6 → K3Fe(CN)6+ e-
Specification
Blood volume | 0.6μL |
Sample type | Capillary whole bloodyenous whole blood |
Calibration | Plasma equivalent |
Measuring time | 5s |
Meter storage/transportation condition | -20℃~55℃ |
Dimension | 84*60*26(mm) |
Weight | 73.5g,with battery |
Power source | 3V DC, 50mA, 2 AAA alkaline batteries |
Memory | 10 blood glucose test results |
Operating condition | 10℃~35℃; ≤80% RH |
Construction | Hand-held |
Measurement units | mg/dL or mmol/L |
Measurement range | 20~600 mg/dL or 1.1~33.3mmol/L |
Shelf life | 10 years(estimated by test 7 times per day). During use, user should maintain the product refer to this user manual’s requirements. |
◆ User performance evaluation:
Results for glucose concentrations<100 mg/dL (<5.55mmol/L) | ||
---|---|---|
Within ± 5mg/dL (within ± 0.28mmol/L) | Within ± 10mg/dL (within ± 0.56mmol/L) | Within ± 15mg/dL (within ± 0.83mmol/L) |
56%(5/9) | 100%(9/9) | 100%(9/9) |
Results for glucose concentrations ≥ 100 mg/dL (≥5.55mmol/L) | ||
Within ± 5% 71%(65/91) | Within ± 10% 93%(85/91) | Within ± 15% 98%(89/91) |