Təhlükəsiz AQ UG
Bi-functional blood glucose and uric acid monitoring system; Automatic strips ejection; ISO15197 2013 approved; Smart code for uric acid test

Baxış
The Safe AQ UG Blood Glucose and Uric Acid Monitoring System is designed for the quantitative measurement of glucose and uric acid in fresh capillary whole blood samples and in venous whole blood samples. The Safe AQ UG Blood Glucose and Uric Acid Monitoring System is for use outside the body only (in vitro diagnostic use) for self-testing and professional use as an aid in the management of diabetes and hyperuricemia (HUA).
spesifikasiya
Test funksiyası | Qan qlükoza | Sidik turşusu |
---|---|---|
test üsulu | FAD-GDH | Oksidləşmiş ferrosen törəməsi |
Nümunə növü | Periferik kapilyar tam qan / venoz tam qan | Periferik kapilyar tam qan / venoz tam qan |
Qan nümunəsi | 0.6μl | 3μl |
Test aralığı | 20-600mg / dL (1.1-33.3mmol / L) | 3.0-20.0mg / dL (181-1188 µmol/L) |
Test vaxtı | 5s | 25s |
Yaddaş | 500 edir | 100 edir |
Şeridin etibarlılığı | 24 ay | 12 ay |
◆ Qan qlükoza istifadəçi performansının qiymətləndirilməsi:
Qlükoza konsentrasiyalarının nəticələri <100 mg / dL (<5.55mmol / l) | ||
---|---|---|
± 5mg / dL daxilində (± 0.28mmol / l daxilində) | ± 10mg / dL daxilində (± 0.56mmol / l daxilində) | ± 15mg / dL daxilində (± 0.83mmol / l daxilində) |
50% (14/28) | 85.71% (24 / 28) | 100% (28 / 28) |
Qlükoza konsentrasiyalarının nəticələri ≥ 100 mg / dL (-5.55mmol / l) | ||
± 5% -də | ± 10% -də | ± 15% -də |
55.56% (40/72) | 90.28% (65/72) | 97.22% (70/72) |
◆ Uric acid user performance evaluation:
Results for uric acid concentrations < 297μmol/L (5.0 mg/ dL) | ||
---|---|---|
Within ± 14.9μmol/L (Within ± 0.3 mg/ dL) | Within ± 29.7μmol/L (Within ± 0.5 mg/ dL) | Within ± 59.4μmol/L (Within ± 1.0 mg/ dL) |
40.0% (6/15) | 60.0% (9 / 15) | 100% (15 / 15) |
Results for uric acid concentrations≥ 297μmol/L (5.0 mg/ dL) | ||
± 5% -də 57.6% (49/85) | ± 10% -də 94.1% (80/85) | ± 15% -də 100% (85/85) |