Sigurno Accu2

Green-Yellow-Red light warning;

Auto sample loading detection;

Auto electrode inserting detection;

Affordable;Code free; CE certification

pregled
  • Namjeravanu upotrebu

    Safe Accu2 is intended for IVD testing the blood glucose in the fresh whole capillary blood and vein plasma; for medical institutes’ fast testing of blood glucose; for diabetics and other groups’ blood glucose monitoring.

  • Princip ispitivanja

    C6H12O6 + K3Fe(CN)6   →   FAD-GDH   →  C6H12O7 + K4Fe(CN)6
                        K4Fe(CN)6 → K3Fe(CN)6+ e-

  • Karakteristike (tehničke prednosti)

    1) FAD-GDH system : It is not easily affected by oxygen molecule in the blood.

    2) User-friendly operation system: no coding & automatic test strip ejectionin the management of diabetes.

specifikacija
Zapremina krvi 0.6μL
Vrsta uzorka Kapilarna cijela krvava cijela krv
kalibrisanje Ekvivalent plazme
Merenje vremena 10s
Stanje skladištenja / transporta brojila -20C ~ 55 ℃
dimenzija 105 * 60 * 23 (mm)
Težina 77.2 g, sa baterijom
Izvor napajanja 3V DC,10mA,2 AAA alkaline batteries
memorija 200 testova glukoze u krvi
Radno stanje Vlažnost 10 ℃ ~ 35C≤80%
izgradnja Ručni
Mjerne jedinice mg / dL ili mmol / L
Merni opseg 20 ~ 600 mg / dL ili 1.1 ~ 33.3 mm / L
Precision: test result of Safe-Accu2 meet the requirement below:

Accuracy: 95% test result of Safe-Accu 2 meet the requirement below:


Raspon koncentracije Pristranost%
<5.5 mmol/L(100 mg/dL) Within±0.83 mmol/L(15 mg/dL)
25.5 mmol/L(100mg/dL) Unutar ± 15%


Precision: test result of Safe-Accu 2 meet the requirement below:


Raspon koncentracije Require ment
<5.5 mmol/L(100 mg/dL) SD<0.34mmol/L(6.0mg/dL)
25.5 mmol/L(100mg/dL) CV <6.0%



                         