Affordable;Code free; CE certification

visió de conjunt
  • Ús previst

    Safe Accu2 is intended for IVD testing the blood glucose in the fresh whole capillary blood and vein plasma; for medical institutes’ fast testing of blood glucose; for diabetics and other groups’ blood glucose monitoring.

  • Principi de prova

    C6H12O6 + K3Fe(CN)6   →   FAD-GDH   →  C6H12O7 + K4Fe(CN)6
                        K4Fe(CN)6 → K3Fe(CN)6+ e-

  • Features (Technical advantages)

    1) FAD-GDH system : It is not easily affected by oxygen molecule in the blood.

    2) User-friendly operation system: no coding & automatic test strip ejectionin the management of diabetes.

Especificació
Volum de sang0.6 μL
Tipus de mostraSang sencera capil·lar
calibratgeEquivalent de plasma
Mesura del temps10s
Condicions d'emmagatzematge/transport del comptador-20C ~ 55℃
dimensió* 105 60 23 * (mm)
pes77.2 g, amb bateria
Font d'energia3V DC,10mA,2 AAA alkaline batteries
Memòria200 resultats de la prova de glucosa en sang
Estat de funcionament10℃~35C≤80% RH
construccióPortat a mà
Unitats de mesuramg / dL o mmol / L
Rang de mesura20~600 mg/dL o 1.1~33.3 mmo/L
Precision: test result of Safe-Accu2 meet the requirement below:

Accuracy: 95% test result of Safe-Accu 2 meet the requirement below:


Rang de concentracióParcialitat%
<5.5 mmol/L(100 mg/dL)Within±0.83 mmol/L(15 mg/dL)
25.5 mmol/L(100mg/dL)Dins del ± 15%


Precision: test result of Safe-Accu 2 meet the requirement below:


Rang de concentracióRequire ment
<5.5 mmol/L(100 mg/dL)SD<0.34mmol/L(6.0mg/dL)
25.5 mmol/L(100mg/dL)CV <6.0%



                         