EN
ĉiuj Kategorioj
EN

Serio Monitoranta Glukozon

Ora AQ Plus

LUNULA nanoTM patent blood sample applying technology; 99.99% pure gold electrode test strip;Widely HCT test range;Strong anti-interference



Superrigardo
  • Intencita uzo

    The Gold AQ Plus Blood Glucose Monitoring system is intended for use in the quantitative measurement of glucose in human whole blood taken from the fingertip (capillary) or from the vein (venous) or from the artery (arterial). It is intended to be used by people with diabetes at home or clinical site as an aid in monitoring the effectiveness of their diabetes control programs.

  • Testoprincipo

    A glucose test is based on the measurement of electrical current caused by the reaction of glucose with the FAD glucose dehydrogenase on the electrode of the strip. The blood or control solution sample is drawn into the tip of the test strip through capillary action. Glucose in the sample reacts with the FAD glucose dehydrogenase and generates electrons, which produce an electrical current. The Gold AQ Plus blood glucose meter measures the electrical current and calculates the glucose result. The glucose result is displayed by the meter in the unit of mg/dL or mmol/L.

  • About the Gold AQ Plus Blood Glucose Test Strip

    The Gold AQ Plus blood glucose test strip is intended to be used with the Gold AQ Plus blood glucose meter to measure the amount of blood glucose in fresh capillary whole blood or venous blood or arterial blood. The Gold AQ Plus blood glucose test strip used with the Gold AQ Plus blood glucose meter provides a complete test system that is intended for in vitro diagnostic use by people with diabetes. The Gold AQ Plus blood glucose test strip do not require coding. Please refer to the “Performing a Blood Glucose Test” section for complete instructions.


The Gold AQ Plus blood glucose test strip consists of the following parts:

specifo
Sanga volumo 0.8μL
Specimen-tipo Tuta sango kapilara
calibración Ekvivalenta plasmo
Mezura tempo 5s
Metra stokado / transporta kondiĉo -20C ~ 55 ℃
dimensio (103mm (L) x 60mm (W) x 24mm(H))
Pezo pri 59g
energifonto 2 AAA alkaline batteries
memoro Rezultaj mezuradoj de 999 sango glukozo
dato kaj horo
Rezultaj mezurrezultoj kun 200 Control Solution
dato kaj horo
Test condition Temperature: 10 ℃ ~ 40 ℃
Relative Humidity: 10% ~ 90% RH (non-condensing)
Hematocrit:20%~70%
Altitude: Up to 10100 feet (3078 meters)
Note: Use within specified environmental conditions only.
Funkcianta kondiĉo 10 ℃ ~ 35C≤80% RH
konstruado Mano-tenado
Mezuraj unuoj mg / dL aŭ mmol / L
Mezura teritorio 20 ~ 600 mg / dL aŭ 1.1 ~ 33.3mmol / L
Prova kemia konsisto FAD glucose dehydrogenase, potassium ferricyanide,
non-reactive ingredients
Control solution chemical composition Water, glucose, preservative, dyes, phosphate buffer,
viscosity enhancing agents
Testo de stokado de strio de provoj 1℃~30℃ 10%~90%RH
breto vivo 10 jaroj (taksita per testo 7 fojojn tage).)
Dum uzo, uzanto devas konservi la produkton
referencu la postulojn de ĉi tiu uzantlibro.


1) 99.99% gold electrode, a conductivity of 1,400 times than carbon electrode.
2) With superior electrochemical property and extremely strong corrosion resistance.
3) Able to detect the blood sample temperature, HCT and ambient temperature accurately, and automatically correct the result deviations caused by above factors.


LUNULA nano patent blood sample apply technology: Upon optimization of the quantity and positions of
air exhausting microporous, the blood applying speed has been obviously improved.