PABA 1000

Best choice for the early screening of renal disease

Sealladh farsaing
  • Cleachdadh air a mholadh

    The PABA-1000 ACR Analyzer is intended to be used with the mALB/Cr Reagent Kit manufactured by Changsha Sinocare Inc. to quantitatively determine microalbuminuria, creatinine and ACR (the ratio of microalbuminuria and creatinine) in urine sample. It is for in vitro diagnostic use only. It is intended for prescription point-of-care use as an aid in the diagnosis of renal disease.

  • -obrach Prionnsapal

    The ACR of tested sample is determined by colorimetry (Cr) and fluorometry (mALB). When passing through tested solution, one light beam is absorbed by Solution A, while another light beam excites the fluorescent substances in Solution B. The content of tested substance, which is clinically significant, can be acquired by analyzing the amount of absorbed light or generated florescence.

  • Liosta pacaidh

    ACR Monitoring System consists of PABA-1000 ACR analyzer, mALB/Cr Reagent Kit and control solutions (level 1, level 2). The mALB/Cr Reagent Kit is intended to be used with the PABA-1000 ACR Analyzer to quantitatively determine microalbuminuria, creatinine and ACR (the ratio of microalbuminuria and creatinine) in urine sample.



Diabetes is the leading cause for the end-stage renal disease


▼In developed countries, about 1/3 of diabetic patients have diabetic kidney disease, which is the main cause of renal failure 1,2

▼According to the DEMAND study, the detection rate of albuminuria in Asian diabetic patients was 56%, higher than that in Caucasian patients (41%) , Early detection of renal dysfunction may arrest/delay the progression of renal failure





5


6


4

Sònrachadh
Notaichean paramadair
Seòrsa sampaill Urine
Reagent

mALB(MAU)/Creatinine test kit

(fluorescence and Benedict-Behre Methods)

cànain
English / simplified Chinese
taisbeanaidh 5-inch touch screen
Communication port RS232,USB,RJ45
Inntrigeadh cumhachd ~220V,50Hz,1.4A
Meud 380mm×240mm×254mm(L×W×H)
cuideam lom
6.5kg
Clò-bhualadair
Clò-bhualadair teirmeach togte
Single test time About 6 min
Data storage volume 2000 test and quality control data
Notaichean

Modheòlas

Raon na Precison

mALB (MAU)

Fluorescence scattering 5.0mg/L-200.0mg/L

5.0~30.0mg/L

SD2.0mg/L;

30.1~200.0mg/L,CV6.5%
Creatinine Benedict-Behre

10.0-400.0mg / dL

(0.88-35.30mmol/L)

10.0-60.0mg/dL SD3.0mg/dL;

60.1-400.0mg/dL CV5%
ACR /

1.3-2000.0mg/g

(0.14-227.28mg/mmol)
CV10%