Green-Yellow-Red light warning; Auto electrode inserting detection; Auto sample loading detection; Affordable;Code free; CE certification

Overview
  • Amfanin da ake nufi

    Safe Accu2 is intended for IVD testing the blood glucose in the fresh whole capillary blood and vein plasma; for medical institutes’ fast testing of blood glucose; for diabetics and other groups’ blood glucose monitoring.

  • Gwajin Gwaji

    C6H12O6 + K3Fe(CN)6 → FAD-GDH → C6H12O7 + K4Fe(CN)6
    K4Fe(CN)6 → K3Fe(CN)6+ e-

  • Fasali (Fa'idodin fasaha)

    1) FAD-GDH system : It is not easily affected by oxygen molecule in the blood.

    2) User-friendly operation system: no coding & automatic test strip ejectionin the management of diabetes.

Musammantawa
Bloodarar jini0.6μL
Nau'in samfuriKayan mulkin mallaka duk jini
kPlasma daidai yake
Lokacin aunawa10s
Mita ajiya / yanayin sufuri-20C ~ 55 ℃
girma105 * 60 * 23 (mm)
Weight77.2g, tare da baturi
ikon source3V DC,10mA,2 AAA alkaline batteries
MemorySakamakon gwajin glucose na jini
Yanayin aiki10 ℃ ~ 35C≤80% RH
ConstructionDa hannu
Na'urar aunawamg / dL ko mmol / L
Tsarin auna20 ~ 600 mg / dL ko 1.1 ~ 33.3mmo / L
Precision: test result of Safe-Accu2 meet the requirement below:

Accuracy: 95% test result of Safe-Accu 2 meet the requirement below:

Rangaran TaroBias%
<5.5 mmol / L (100 mg / dL)Within±0.83 mmol/L(15 mg/dL)
25.5 mmol / L (100mg / dL)A tsakanin ± 15%

Precision: test result of Safe-Accu 2 meet the requirement below:

Rangaran TaroNa bukatar ment
<5.5 mmol / L (100 mg / dL)SD<0.34mmol/L(6.0mg/dL)
25.5 mmol / L (100mg / dL)CV <6.0%