Hoʻoikaika i ka Pūnaehana Puʻuwai koko

Kāhea paʻa ʻo Accu2

Green-Yellow-Red light warning; Auto electrode inserting detection; Auto sample loading detection; Affordable;Code free; CE certification

Overview
  • Hoʻohana hoʻohana

    Safe Accu2 is intended for IVD testing the blood glucose in the fresh whole capillary blood and vein plasma; for medical institutes’ fast testing of blood glucose; for diabetics and other groups’ blood glucose monitoring.

  • Test Principle

    C6H12O6 + K3Fe(CN)6 → FAD-GDH → C6H12O7 + K4Fe(CN)6
    K4Fe(CN)6 → K3Fe(CN)6+ e-

  • Features (Technical advantages)

    1) FAD-GDH system : It is not easily affected by oxygen molecule in the blood.

    2) User-friendly operation system: no coding & automatic test strip ejectionin the management of diabetes.

hoakaka
Nā nui koko0.6μL
ʻApana ʻanoKe kō koko piha koko nui
Ka hoʻopiliʻanaPalapala hoʻohālike
Waena ʻia10s
Ke waiho ʻana i ke ʻano no ka mālama ʻana / me ke ala-20C~55℃
Ana105*60*23(mm)
Weight77.2g,with battery
Power kumu3V DC,10mA,2 AAA alkaline batteries
hoomanao ana200 nā hopena hoʻāʻo glucose koko
Ke ʻano hana10℃~35C≤80% RH
ieoaeunoaiPolika paʻa lima
Nā kikowaena kikowaenamg / dL or mmol / L
Ka pae ana20 ~ 600 mg / dL or 1.1 ~ 33.3mmo / L
Precision: test result of Safe-Accu2 meet the requirement below:

Accuracy: 95% test result of Safe-Accu 2 meet the requirement below:

Palapala UloBias%
<5.5 mmol / L (100 mg / dL)Within±0.83 mmol/L(15 mg/dL)
25.5 mmol / L (100mg / dL)Within ± 15%

Precision: test result of Safe-Accu 2 meet the requirement below:

Palapala UloMakemake ment
<5.5 mmol / L (100 mg / dL)SD<0.34mmol/L(6.0mg/dL)
25.5 mmol / L (100mg / dL)CV <6.0%