Kāhea paʻa ʻo Accu2
Green-Yellow-Red light warning; Auto electrode inserting detection; Auto sample loading detection; Affordable;Code free; CE certification

Overview
Hoʻohana hoʻohana
Safe Accu2 is intended for IVD testing the blood glucose in the fresh whole capillary blood and vein plasma; for medical institutes’ fast testing of blood glucose; for diabetics and other groups’ blood glucose monitoring.
Test Principle
C6H12O6 + K3Fe(CN)6 → FAD-GDH → C6H12O7 + K4Fe(CN)6
K4Fe(CN)6 → K3Fe(CN)6+ e-Features (Technical advantages)
1) FAD-GDH system : It is not easily affected by oxygen molecule in the blood.
2) User-friendly operation system: no coding & automatic test strip ejectionin the management of diabetes.
hoakaka
Nā nui koko | 0.6μL |
ʻApana ʻano | Ke kō koko piha koko nui |
Ka hoʻopiliʻana | Palapala hoʻohālike |
Waena ʻia | 10s |
Ke waiho ʻana i ke ʻano no ka mālama ʻana / me ke ala | -20C~55℃ |
Ana | 105*60*23(mm) |
Weight | 77.2g,with battery |
Power kumu | 3V DC,10mA,2 AAA alkaline batteries |
hoomanao ana | 200 nā hopena hoʻāʻo glucose koko |
Ke ʻano hana | 10℃~35C≤80% RH |
ieoaeunoai | Polika paʻa lima |
Nā kikowaena kikowaena | mg / dL or mmol / L |
Ka pae ana | 20 ~ 600 mg / dL or 1.1 ~ 33.3mmo / L |
Precision: test result of Safe-Accu2 meet the requirement below:
Accuracy: 95% test result of Safe-Accu 2 meet the requirement below:
Palapala Ulo | Bias% |
---|---|
<5.5 mmol / L (100 mg / dL) | Within±0.83 mmol/L(15 mg/dL) |
25.5 mmol / L (100mg / dL) | Within ± 15% |
Precision: test result of Safe-Accu 2 meet the requirement below:
Palapala Ulo | Makemake ment |
---|---|
<5.5 mmol / L (100 mg / dL) | SD<0.34mmol/L(6.0mg/dL) |
25.5 mmol / L (100mg / dL) | CV <6.0% |