Hoʻoikaika i ka Pūnaehana Puʻuwai koko

ʻOi mau ka AQ Smart

Fast 5s test time; FDA-GDH monitoring system; SINODRAW patent lancing device; Temperature warning

Overview
  • Overview

    The Safe AQ Smart Blood Glucose Monitoring System is designed for the quantitative measurement of glucose in fresh capillary whole blood samples taken from the fingertip and in venous whole blood samples. The Blood Glucose Monitoring System is for use outside the body only (in vitro diagnostic use) for self-testing and professional use as an aid in the management of diabetes.

  • Test principle

    A blood glucose test is based on measurement of electrical current caused by the reaction of glucose with the reagents (special chemicals) on the electrode of the strip. The blood or control solution sample is drawn into the tip of the test strip through capillary action. Glucose in the sample reacts with the special chemicals and generates electrons, which produce electrical current. The blood glucose meter measures the electrical current and calculates the glucose result. The glucose result is displayed by your meter as mg/dL or mmol/L.

  • Nā hiʻohiʻona (Nā pono kūpono)

    1) Pūnaehana FAD-GDH: ʻAʻole ia i ka maʻalahi i ka molekle oxygen i loko o ke koko.

    2) Pūnaehana hana lawelawe aloha a ka mea hoʻohana: ʻaʻohe coding & nā uila hoʻokolohua kūlohelohe ejectionin e hoʻokele i ka maʻi maʻi.

hoakaka
Nā nui koko0.6μL
ʻApana ʻanoKe kō koko piha koko nui
Ka hoʻopiliʻanaPalapala hoʻohālike
Waena ʻia5s
Ke waiho ʻana i ke ʻano no ka mālama ʻana / me ke ala-20~ 55 ℃
Ana84 * 60 * 26 (mm)
Weight73.5g, with battery
Power kumu3V DC, 50mA, 2 AAA alkaline batteries
hoomanao ana10 nā hopena hoʻāʻo glucose koko
Ke ʻano hana10 ℃ ~ 35; ≤80% RH
ieoaeunoaiPolika paʻa lima
Nā kikowaena kikowaenamg / dL or mmol / L
Ka pae ana20 ~ 600 mg / dL or 1.1 ~ 33.3mmo / L
Ola ola10 mau makahiki (kuhi ʻia ma ka hōʻike ʻo 7 mau manawa i kēlā me kēia lā).
I ka wā hoʻohana, pono ka mea hoʻohana e mālama i ka huahana
e pili i nā pono o ka hoʻohanaʻana i kēia mea hoʻohana.
Nā Uihana Hana

Accuracy: 95% test result of Safe AQ meet the requirement below:

Palapala UloBias%
<5.5 mmol / L (100 mg / dL)I loko o ± 0.83 mmol / L (15 mg / dL)
25.5 mmol / L (100mg / dL)I loko o ± 15%

Precision: test result of Safe AQ Smart meet the requirement below:

Palapala UloMakemake ment
<5.5 mmol / L (100 mg / dL)SD <0.34mmol / L (6.0mg / dL)
25.5 mmol / L (100mg / dL)CV <6.0%