ʻOi mau ka AQ Smart
Fast 5s test time; FDA-GDH monitoring system; SINODRAW patent lancing device; Temperature warning

Overview
Overview
The Safe AQ Smart Blood Glucose Monitoring System is designed for the quantitative measurement of glucose in fresh capillary whole blood samples taken from the fingertip and in venous whole blood samples. The Blood Glucose Monitoring System is for use outside the body only (in vitro diagnostic use) for self-testing and professional use as an aid in the management of diabetes.
Test principle
A blood glucose test is based on measurement of electrical current caused by the reaction of glucose with the reagents (special chemicals) on the electrode of the strip. The blood or control solution sample is drawn into the tip of the test strip through capillary action. Glucose in the sample reacts with the special chemicals and generates electrons, which produce electrical current. The blood glucose meter measures the electrical current and calculates the glucose result. The glucose result is displayed by your meter as mg/dL or mmol/L.
Nā hiʻohiʻona (Nā pono kūpono)
1) Pūnaehana FAD-GDH: ʻAʻole ia i ka maʻalahi i ka molekle oxygen i loko o ke koko.
2) Pūnaehana hana lawelawe aloha a ka mea hoʻohana: ʻaʻohe coding & nā uila hoʻokolohua kūlohelohe ejectionin e hoʻokele i ka maʻi maʻi.
hoakaka
Nā nui koko | 0.6μL |
ʻApana ʻano | Ke kō koko piha koko nui |
Ka hoʻopiliʻana | Palapala hoʻohālike |
Waena ʻia | 5s |
Ke waiho ʻana i ke ʻano no ka mālama ʻana / me ke ala | -20℃~ 55 ℃ |
Ana | 84 * 60 * 26 (mm) |
Weight | 73.5g, with battery |
Power kumu | 3V DC, 50mA, 2 AAA alkaline batteries |
hoomanao ana | 10 nā hopena hoʻāʻo glucose koko |
Ke ʻano hana | 10 ℃ ~ 35℃; ≤80% RH |
ieoaeunoai | Polika paʻa lima |
Nā kikowaena kikowaena | mg / dL or mmol / L |
Ka pae ana | 20 ~ 600 mg / dL or 1.1 ~ 33.3mmo / L |
Ola ola | 10 mau makahiki (kuhi ʻia ma ka hōʻike ʻo 7 mau manawa i kēlā me kēia lā). I ka wā hoʻohana, pono ka mea hoʻohana e mālama i ka huahana e pili i nā pono o ka hoʻohanaʻana i kēia mea hoʻohana. |
Nā Uihana Hana
Accuracy: 95% test result of Safe AQ meet the requirement below:
Palapala Ulo | Bias% |
---|---|
<5.5 mmol / L (100 mg / dL) | I loko o ± 0.83 mmol / L (15 mg / dL) |
25.5 mmol / L (100mg / dL) | I loko o ± 15% |
Precision: test result of Safe AQ Smart meet the requirement below:
Palapala Ulo | Makemake ment |
---|---|
<5.5 mmol / L (100 mg / dL) | SD <0.34mmol / L (6.0mg / dL) |
25.5 mmol / L (100mg / dL) | CV <6.0% |