EN
Txhua pawg
EN

Ntshav Qab Zib Ntsuam Xyuas Series

Kev Nyab Xeeb Accu2

Green-Yellow-Red light warning; Auto electrode inserting detection; Auto sample loading detection; Affordable;Code free; CE certification

Txheej txheem cej luam
  • Siv Tau

    Safe Accu2 is intended for IVD testing the blood glucose in the fresh whole capillary blood and vein plasma; for medical institutes’ fast testing of blood glucose; for diabetics and other groups’ blood glucose monitoring.

  • Xeem Ntawv Tso Cai

    C6H12O6 + K3Fe(CN)6 → FAD-GDH → C6H12O7 + K4Fe(CN)6
    K4Fe(CN)6 → K3Fe(CN)6+ e-

  • Nta (Kev zoo)

    1) FAD-GDH system: Nws tsis yooj yim cuam tshuam los ntawm oxygen molecule hauv cov ntshav.

    2) Kev siv txoj haujlwm siv rau cov neeg siv khoom: tsis muaj coding & tsis siv neeg ntsuas cov kab tawm ejectionin tswj kev mob ntshav qab zib.

specification
Ntshav ntim0.6μL
Qauv homCapillary tag nrho cov ntshav ntshav tag nrho cov ntshav
calibrationNtshav sib npaug
Kev ntsuas lub sijhawm10s
Meter cia / tsheb thauj mus los mob-20C ~ 55 ℃
dimension105 * 60 * 23 (mm)
Nyhav77.2g,with battery
hwj chim qhov twg los3V DC,10mA,2 AAA alkaline batteries
Memory200 cov ntshav kuaj cov ntshav
Kev ua haujlwm ntawm kev ua haujlwm10 ℃ ~ 35C≤80% RH
siv cov kev faibTes tuav
Kev ntsuas ntsuasmg / dL lossis mmol / L
Kev ntsuas ntau20 ~ 600 mg / dL lossis 1.1 ~ 33.3mmo / L
Precision: test result of Safe-Accu2 meet the requirement below:

Accuracy: 95% test result of Safe-Accu 2 meet the requirement below:

Kev Ntsiag To Ntau YamKev tsis ncaj ncees%
<5.5 mmol / L (100 mg / dL)Nyob hauv ± 0.83 mmol / L (15 mg / dL)
25.5 mmol / L (100mg / dL)Nyob rau hauv ± 15%

Precision: test result of Safe-Accu 2 meet the requirement below:

Kev Ntsiag To Ntau YamYuav tsum tau ment
<5.5 mmol / L (100 mg / dL)SD <0.34mmol / L (6.0mg / dL)
25.5 mmol / L (100mg / dL)CV <6.0%