Fast 5s test time & Patent painless lancing device & Non-interference & Voice reminder & Red-green light warning.

Txheej txheem cej luam

Safe AQ Voice Blood Glucose Monitoring System is intended for in vitro diagnostic use and should not be used for the diagnosis of or screening of diabetes.

  • pob

    The Safe AQ Voice Blood Glucose Monitoring System includes:
    Safe AQ Voice blood glucose meter, Safe AQ blood glucose test strip, blood glucose control solution

  • Nta (Kev zoo)

    1) FAD-GDH system: It is not easily affected by oxygen molecule in the blood.
    2) User-friendly operation system: fast 5s test time & patent painless lancing device & non-interference & voice reminder & red-green light warning.

  • Kev ntsuas cov qauv

    C6H12O6 + K3Fe(CN)6 → FAD-GDH → C6H12O7 + K4Fe(CN)6
    K4Fe(CN)6 → K3Fe(CN)6+ e-

specification
Ntshav ntim0.6μL
Qauv homCapillary tag nrho cov ntshav ntshav tag nrho cov ntshav
calibrationNtshav sib npaug
Kev ntsuas lub sijhawm5s
Meter cia / tsheb thauj mus los mob-20~ 55 ℃
dimension84 * 60 * 26 (mm)
Nyhav73.5g, nrog roj teeb
hwj chim qhov twg los3V DC, 50mA, 2 AAA roj teeb lub roj teeb
Memory10 cov ntshav kuaj cov ntshav
Kev ua haujlwm ntawm kev ua haujlwm10 ℃ ~ 35; ≤80% RH
siv cov kev faibTes tuav
Kev ntsuas ntsuasmg / dL lossis mmol / L
Kev ntsuas ntau20 ~ 600 mg / dL lossis 1.1 ~ 33.3mmol / L
txee lub neej10 xyoo (kwv yees ntsuas ntawm 7 zaug nyob rau ib hnub).
Thaum siv, neeg siv yuav tsum tswj cov khoom lag luam
saib mus rau phau ntawv siv siv ntawm no cov cai.
Evaluation Kev ntsuas cov neeg siv khoom:

Results for glucose concentrations<100 mg/dL (<5.55mmol/L)
Nyob rau hauv ± 5mg / dL
(within ± 0.28mmol/L)
Nyob rau hauv ± 10mg / dL
(within ± 0.56mmol/L)
Nyob rau hauv ± 15mg / dL
(within ± 0.83mmol/L)
56% (5/9)100% (9/9)100% (9/9)
Results for glucose concentrations ≥ 100 mg/dL (≥5.55mmol/L)
Nyob rau hauv ± 5%
71% (65/91)
Nyob rau hauv ± 10%
93% (85/91)
Nyob rau hauv ± 15%
98% (89/91)