Fast 5s test time; FDA-GDH monitoring system; SINODRAW patent lancing device; Temperature warning

Pregled
  • Pregled

    The Safe AQ Smart Blood Glucose Monitoring System is designed for the quantitative measurement of glucose in fresh capillary whole blood samples taken from the fingertip and in venous whole blood samples. The Blood Glucose Monitoring System is for use outside the body only (in vitro diagnostic use) for self-testing and professional use as an aid in the management of diabetes.

  • Test principle

    A blood glucose test is based on measurement of electrical current caused by the reaction of glucose with the reagents (special chemicals) on the electrode of the strip. The blood or control solution sample is drawn into the tip of the test strip through capillary action. Glucose in the sample reacts with the special chemicals and generates electrons, which produce electrical current. The blood glucose meter measures the electrical current and calculates the glucose result. The glucose result is displayed by your meter as mg/dL or mmol/L.

  • Features (Technical advantages)

    1) FAD-GDH system : It is not easily affected by oxygen molecule in the blood.

    2) User-friendly operation system: no coding & automatic test strip ejectionin the management of diabetes.

Specifikacija
Volumen krvi 0.6μL
Vrsta uzorka Kapilarna cjelovita krv
Kalibracija Ekvivalent u plazmi
Mjerenje vremena 5s
Stanje skladištenja / prijevoza brojila -20~55℃
Dimenzija 84 * 60 * 26 (mm)
Težina 73.5g, with battery
Izvor napajanja 3V DC, 50mA, 2 AAA alkaline batteries
memorija 10 blood glucose test results
Radni uvjet 10℃~35; ≤80% RH
Gradnja Ručna
Mjerne jedinice mg / dL ili mmol / L
Raspon mjerenja 20~600 mg/dL or 1.1~33.3mmo/L
Rok trajanja 10 godina (procjenjuje se testnim 7 puta dnevno).
Tijekom upotrebe, korisnik bi trebao održavati proizvod
pogledajte zahtjeve ovog korisničkog priručnika.
Karakteristike izvedbe

Accuracy: 95% test result of Safe AQ meet the requirement below:

Concentration Range Bias%
<5.5 mmol/L(100 mg/dL) Within±0.83 mmol/L(15 mg/dL)
25.5 mmol/L(100mg/dL) Unutar ± 15%

Precision: test result of Safe AQ Smart meet the requirement below:

Concentration Range Require ment
<5.5 mmol/L(100 mg/dL) SD<0.34mmol/L(6.0mg/dL)
25.5 mmol/L(100mg/dL) CV<6.0%