Seri Pemantauan Glukosa Darah

Accu Aman

FAD-GDH system: It is not easily affected by oxygen molecule in the blood; User-friendly operation system: no coding & automatic test strip ejection


C6H12O6 + K3Fe(CN)6 FAD-GDH C6H12O7 + K4Fe(CN)6

Gambaran Umum
  • Deskripsi

    The Safe-Accu Blood Glucose Monitoring System includes: Safe-Accu blood glucose meter, Safe-Accu blood glucose test strip, blood glucose control solution PLEASE NOTE: The blood glucose control solution is optional.

  • Aplikasi

    The Safe-Accu Blood Glucose Monitoring System is designed for the quantitative measurement of glucose in fresh capillary whole blood samples taken from the fingertip and in venous whole blood samples. The Blood Glucose Monitoring System is for use outside the body only (in vitro diagnostic use) for self-testing and professional use as an aid in the management of diabetes.

  • Penggunaan yang dimaksudkan

    Safe Accu blood glucose monitoring system is intended for external use only and widely applied in monitoring blood glucose. It is designed to be easily operating, convenient and only need little volume of blood sample. All these advantages make it an excellent blood glucose monitoring tool.

Spesifikasi
Volume darah 0.6μL
Jenis sampel Seluruh darah kapiler berdarah lengkap
Kalibrasi Setara plasma
Mengukur waktu 10±1 seconds
Penyimpanan meter / kondisi transportasi -20~ 55 ℃
Dimensi 96 * 55 * 18 (mm)
Berat tentang40g
Sumber daya 3V DC(CR2032) button cell
Battery service time Support 1000 times testing
Ingatan 200 hasil tes glukosa darah
Kondisi operasi 10 ℃ ~ 35; ≤80% RH
Konstruksi Genggam
Unit pengukuran mg / dL atau mmol / L
Jarak pengukuran 20 ~ 600 mg / dL atau 1.1 ~ 33.3mmol / L
SoftwareVersion 20131120
Precision: test result of Safe-Accu meet the requirement below:
Rentang Konsentrasi Membutuhkan ment
<5.5 mmol / L (100 mg / dL) SD<0.34mmol(6.0mg/dL)
25.5 mmol / L (100mg / dL) CV <6.0%
More details please read Safe-Accu blood glucose test strip package insert.