Bi-functional blood glucose and uric acid monitoring system; Automatic strips ejection; ISO15197 2013 approved; Smart code for uric acid test

Overview

The Safe AQ UG Blood Glucose and Uric Acid Monitoring System is designed for the quantitative measurement of glucose and uric acid in fresh capillary whole blood samples and in venous whole blood samples. The Safe AQ UG Blood Glucose and Uric Acid Monitoring System is for use outside the body only (in vitro diagnostic use) for self-testing and professional use as an aid in the management of diabetes and hyperuricemia (HUA).

nkọwapụta
Ọrụ NleleỌbara glucoseUric Acid
Ule nnwaleFAD-GDHOxidized Ferrocene Derivative
Samdị nleleOria ahihia mmadu nile di ya na aru mmadu / ebe ndi mmadu dumOria ahihia mmadu nile di ya na aru mmadu / ebe ndi mmadu dum
Nyocha ọbara
0.6µl3µl
Ule ule20-600mg / dL
(1.1-33.3mmol / L)
3.0-20.0mg / dL
(181-1188 µmol/L)
Oge nnwale5s25s
na ebe nchekwa500 tent100 tent
Eriri ndaba24 ọnwa17 ọnwa
Evaluation Nyocha onye ọrụ glucose ọbara:
Nsonaazụ maka nnabata glucose <100 mg / dL (<5.55mmol / l)
N'ime ± 5mg / dL
(n'ime ± 0.28mmol / l)
N'ime ± 10mg / dL
(n'ime ± 0.56mmol / l)
N'ime ± 15mg / dL
(n'ime ± 0.83mmol / l)
50% (14/28)85.71% (24/28)100% (28/28)
Nsonaazụ maka nchịkọta glucose ≥ 100 mg / dL (≥5.55mmol / l)
N'ime ± 5%
N'ime ± 10%
N'ime ± 15%
55.56% (40/72)90.28% (65/72)97.22% (70/72)
◆ Uric acid user performance evaluation:
Results for uric acid concentrations < 297μmol/L (5.0 mg/ dL)
Within ± 14.9μmol/L
(Within ± 0.3 mg/ dL)
Within ± 29.7μmol/L
(Within ± 0.5 mg/ dL)
Within ± 59.4μmol/L
(Within ± 1.0 mg/ dL)
40.0% (6/15)60.0% (9/15)100% (15/15)
Results for uric acid concentrations≥ 297μmol/L (5.0 mg/ dL)
N'ime ± 5%
57.6% (49/85)
N'ime ± 10%
94.1% (80/85)
N'ime ± 15%
100% (85/85)