Fast 5s test time & Patent painless lancing device & Non-interference & Voice reminder & Red-green light warning.

Overview
Safe AQ Voice Blood Glucose Monitoring System is intended for in vitro diagnostic use and should not be used for the diagnosis of or screening of diabetes.
-
sarcina
The Safe AQ Voice Blood Glucose Monitoring System includes:
Safe AQ Voice blood glucose meter, Safe AQ blood glucose test strip, blood glucose control solution -
Features (Technical commoda)
1) FAD-GDH system: It is not easily affected by oxygen molecule in the blood.
2) User-friendly operation system: fast 5s test time & patent painless lancing device & non-interference & voice reminder & red-green light warning. -
test principle
C6H12O6 + K3Fe(CN)6 → FAD-GDH → C6H12O7 + K4Fe(CN)6
K4Fe(CN)6 → K3Fe(CN)6+ e-
specification
sanguis volumine | 0.6μL |
Sample genus | Capillaribus totum bloodyenous sanguinis |
Calibration | plasma equivalent |
tunc mensuræ | 5s |
Meter repono / translationem conditione, | -20℃55 ℃ ~ |
ratio | LXXXIV * * LX XXVI (mm) |
Pondus | 73.5g, cum altilium |
Potentia Source | 3v DC, 50mA, II alkaline AAA batteries |
Memoria | 10 sanguinem GLYCOSA test results |
Operating conditione, | 10 ℃ ~ 35℃; ≤80% RH |
Construction | Manibus tenuit, |
unitates mensurae | mg / g, sive DL / L |
measurement range | 20 ~ 600 mg / ~ 1.1mmol 33.3 sive DL / L |
Fasciae vita | 10 annis (taxata test 7 temporibus per diem). Per usus productum in user oportet ponere decurritur ad user manual est scriptor accommodabat. |
◆ User perficientur iudicium:
Results for glucose concentrations<100 mg/dL (<5.55mmol/L) | ||
---|---|---|
In 5mg ± / DL (within ± 0.28mmol/L) |
In 10mg ± / DL (within ± 0.56mmol/L) |
In 15mg ± / DL (within ± 0.83mmol/L) |
56% (5 / 9) | 100% (9 / 9) | 100% (9 / 9) |
Results for glucose concentrations ≥ 100 mg/dL (≥5.55mmol/L) | ||
± in 5% 71% (65 / 91) |
± in 10% 93% (85 / 91) |
± in 15% 98% (89 / 91) |