Accu2 ປອດໄພ

Green-Yellow-Red light warning;

Auto sample loading detection;

Auto electrode inserting detection;

Affordable;Code free; CE certification

ພາບລວມ
  • ການນໍາໃຊ້ທີ່ຖືກຕ້ອງ

    Safe Accu2 is intended for IVD testing the blood glucose in the fresh whole capillary blood and vein plasma; for medical institutes’ fast testing of blood glucose; for diabetics and other groups’ blood glucose monitoring.

  • ຫຼັກການທົດສອບ

    C6H12O6 + K3Fe(CN)6   →   FAD-GDH   →  C6H12O7 + K4Fe(CN)6
                        K4Fe(CN)6 → K3Fe(CN)6+ e-

  • Features (Technical advantages)

    1) FAD-GDH system : It is not easily affected by oxygen molecule in the blood.

    2) User-friendly operation system: no coding & automatic test strip ejectionin the management of diabetes.

ຂໍ້ມູນ
ປະລິມານເລືອດ0.6 ລີດ
ປະເພດຕົວຢ່າງCapillary ເລືອດທັງຫມົດເລືອດ
Calibrationທຽບເທົ່າ Plasma
ການວັດແທກເວລາ10
ການເກັບຮັກສາແມັດ / ສະພາບການຂົນສົ່ງ-20C~55℃
ຂະຫນາດ105 * 60 * 23 (mm)
ນ້ໍາ77.2g, ພ້ອມແບັດເຕີລີ
ແຫຼ່ງ​ພະ​ລັງ​ງານ3V DC, 10mA, 2 AAA alkaline ຫມໍ້ໄຟ
ຫນ່ວຍຄວາມຈໍາ200 ຜົນ​ການ​ທົດ​ສອບ​ນ​້​ໍ​າ​ໃນ​ເລືອດ​
ເງື່ອນໄຂການເຮັດວຽກ10℃~35C≤80% RH
ການກໍ່ສ້າງເຮັດດ້ວຍມື
ຫົວ ໜ່ວຍ ວັດແທກmg / dL ຫຼື mmol / L
ລະດັບການວັດແທກ20~600 mg/dL ຫຼື 1.1~33.3mmo/L
Precision: test result of Safe-Accu2 meet the requirement below:

Accuracy: 95% test result of Safe-Accu 2 meet the requirement below:


ຊ່ວງຄວາມເຂັ້ມຂົ້ນອະຄະຕິ
<5.5 mmol/L(100 mg/dL)Within±0.83 mmol/L(15 mg/dL)
25.5 mmol/L(100mg/dL)ພາຍໃນ± 15%


Precision: test result of Safe-Accu 2 meet the requirement below:


ຊ່ວງຄວາມເຂັ້ມຂົ້ນRequire ment
<5.5 mmol/L(100 mg/dL)SD<0.34mmol/L(6.0mg/dL)
25.5 mmol/L(100mg/dL)CV <6.0%



                         