Safe Accu2

Green-Yellow-Red light warning;

Auto sample loading detection;

Auto electrode inserting detection;

Affordable;Code free; CE certification

Overview
  • Fampiasana tena ilaina

    Safe Accu2 is intended for IVD testing the blood glucose in the fresh whole capillary blood and vein plasma; for medical institutes’ fast testing of blood glucose; for diabetics and other groups’ blood glucose monitoring.

  • Fitsipi-pitsapana

    C6H12O6 + K3Fe(CN)6   →   FAD-GDH   →  C6H12O7 + K4Fe(CN)6
                        K4Fe(CN)6 → K3Fe(CN)6+ e-

  • Toetra (Tombony ara-teknika)

    1) FAD-GDH system : It is not easily affected by oxygen molecule in the blood.

    2) User-friendly operation system: no coding & automatic test strip ejectionin the management of diabetes.

famaritana
Volan'ny rà0.6μL
Karazana santionanyRa manontolo feno rà mandriaka
CalibrationMitovy amin'ny Plasma
Fotoana fandrefesana10s
Toetran'ny fitahirizana / fitaterana metatra-20C~55℃
lafiny105 * 60 * 23 (MG)
Weight77.2g, misy bateria
Power loharano3V DC,10mA,2 AAA alkaline batteries
Memory200 ny valin'ny fitiliana glucose amin'ny ra
Toe-javatra miasa10 ℃ ~ 35C ≤ 80% RH
ConstructionHand-natao
Fitaovana fandrefesanamg / dL na mmol / L
Dingana fametrahana20 ~ 600 mg / dL na 1.1 ~ 33.3mmo / L
Precision: test result of Safe-Accu2 meet the requirement below:

Accuracy: 95% test result of Safe-Accu 2 meet the requirement below:


Sehatra fifantohana%
<5.5 mmol/L(100 mg/dL)Within±0.83 mmol/L(15 mg/dL)
25.5 mmol/L(100mg/dL)Tao anatin'ny ± 15%


Precision: test result of Safe-Accu 2 meet the requirement below:


Sehatra fifantohanaMitaky ment
<5.5 mmol/L(100 mg/dL)SD<0.34mmol/L(6.0mg/dL)
25.5 mmol/L(100mg/dL)CV <6.0%



                         