Smart AQ Smart
Fast 5s test time; FDA-GDH monitoring system; SINODRAW patent lancing device; Temperature warning

Overview
-
Overview
The Safe AQ Smart Blood Glucose Monitoring System is designed for the quantitative measurement of glucose in fresh capillary whole blood samples taken from the fingertip and in venous whole blood samples. The Blood Glucose Monitoring System is for use outside the body only (in vitro diagnostic use) for self-testing and professional use as an aid in the management of diabetes.
-
Test principle
A blood glucose test is based on measurement of electrical current caused by the reaction of glucose with the reagents (special chemicals) on the electrode of the strip. The blood or control solution sample is drawn into the tip of the test strip through capillary action. Glucose in the sample reacts with the special chemicals and generates electrons, which produce electrical current. The blood glucose meter measures the electrical current and calculates the glucose result. The glucose result is displayed by your meter as mg/dL or mmol/L.
-
Endri-javatra (tombony ara-teknika)
1) Rafitra FAD-GDH: Tsy mora voan'ny molekiola oksizenina ao anaty ra izy.
2) rafi-piasana mpampiasa mora vidy: tsy misy coding & fitsapana fanatanjahan-tena otomatis fitantanana ny diabeta.
famaritana
Hatran'ny ra | 0.6μL |
Karazana santionany | Aretina mihoson-dra tanteraka |
Calibration | Mitovy amin'ny Plasma |
Fandrefesana fotoana | 5s |
Toerana fitehirizana metaly / fitaterana | -20℃~ 55 ℃ |
lafiny | 84 * 60 * 26 (MG) |
Weight | 73.5g, with battery |
Power loharano | 3V DC, 50mA, 2 AAA alkaline batteries |
Memory | 10 valin'ny fitiliana gliôma |
Fomba fandidiana | 10 ℃ ~ 35℃; ≤80% RH |
Construction | Hand-natao |
Vatoka fandrefesana | mg / dL na mmol / L |
Dingana fametrahana | 20 ~ 600 mg / dL na 1.1 ~ 33.3mmo / L |
Ny fiainako | 10 taona (tombanana amin'ny fotoana fitsapana 7 isan'andro isan'andro). Mandritra ny fampiasana, ny mpampiasa dia tokony hitazona ny vokatra Jereo ireo zavatra takian'ny mpampiasa ity. |
Performance characteristics
Accuracy: 95% test result of Safe AQ meet the requirement below:
Sehatra fifantohana | Bias% |
---|---|
<5.5 mmol / L (100 mg / dL) | Tao anatin'ny ± 0.83 mmol / L (15 mg / dL) |
25.5 mmol / L (100mg / dL) | Tao anatin'ny ± 15% |
Precision: test result of Safe AQ Smart meet the requirement below:
Sehatra fifantohana | Mila fangatahana |
---|---|
<5.5 mmol / L (100 mg / dL) | SD <0.34mmol / L (6.0mg / dl) |
25.5 mmol / L (100mg / dL) | CV <6.0% |