Safe Accu2

Green-Yellow-Red light warning;

Auto sample loading detection;

Auto electrode inserting detection;

Affordable;Code free; CE certification

Oversikt
  • Tiltenkt bruk

    Safe Accu2 is intended for IVD testing the blood glucose in the fresh whole capillary blood and vein plasma; for medical institutes’ fast testing of blood glucose; for diabetics and other groups’ blood glucose monitoring.

  • Testprinsipp

    C6H12O6 + K3Fe(CN)6   →   FAD-GDH   →  C6H12O7 + K4Fe(CN)6
                        K4Fe(CN)6 → K3Fe(CN)6+ e-

  • Funksjoner (tekniske fordeler)

    1) FAD-GDH system : It is not easily affected by oxygen molecule in the blood.

    2) User-friendly operation system: no coding & automatic test strip ejectionin the management of diabetes.

Spesifikasjon
Blodvolum0.6 μL
PrøvenypeKapillært helblodig helblod
KalibreringPlasmakvivalent
Målingstid10s
Målerlagring / transporttilstand-20C ~ 55 ℃
Dimensjon105 * 60 * 23 (mm)
Vekt77.2g,with battery
Strømkilde3V DC,10mA,2 AAA alkaline batteries
Minne200 blodsukkerresultater
Driftstilstand10 ℃ ~ 35C≤80% RF
AnleggHånd
Måleenhetermg / dL eller mmol / l
Måleområde20 ~ 600 mg / dL eller 1.1 ~ 33.3 mmo / L
Precision: test result of Safe-Accu2 meet the requirement below:

Accuracy: 95% test result of Safe-Accu 2 meet the requirement below:


KonsentrasjonsområdePartiskhet%
<5.5 mmol / L (100 mg / dL)Within±0.83 mmol/L(15 mg/dL)
25.5 mmol / L (100 mg / dL)Innen ± 15%


Precision: test result of Safe-Accu 2 meet the requirement below:


KonsentrasjonsområdeKrav
<5.5 mmol / L (100 mg / dL)SD<0.34mmol/L(6.0mg/dL)
25.5 mmol / L (100 mg / dL)CV <6.0%



                         