EN
Categories All
EN

Lenane la Tlhahlobo ea Mekha ea Mali

Lentsoe le Sireletsehileng la AQ

Fast 5s test time & Patent painless lancing device & Non-interference & Voice reminder & Red-green light warning.

Overview

Safe AQ Voice Blood Glucose Monitoring System is intended for in vitro diagnostic use and should not be used for the diagnosis of or screening of diabetes.

  • sephutheloana

    The Safe AQ Voice Blood Glucose Monitoring System includes:
    Safe AQ Voice blood glucose meter, Safe AQ blood glucose test strip, blood glucose control solution

  • Likarolo (Melemo ea botekgeniki)

    1) FAD-GDH system: It is not easily affected by oxygen molecule in the blood.
    2) User-friendly operation system: fast 5s test time & patent painless lancing device & non-interference & voice reminder & red-green light warning.

  • Test principle

    C6H12O6 + K3Fe(CN)6 → FAD-GDH → C6H12O7 + K4Fe(CN)6
    K4Fe(CN)6 → K3Fe(CN)6+ e-

ga tlhaloso
Bophahamo ba mali0.6μL
Mohlala oa mohlala'Mele oa mali kaofela
PhaputsoPlasma e lekanang
Nako ea ho lekanya5s
Boemo ba ho boloka limmapa / boemelo-20~ 55 ℃
teba ka84 * 60 * 26 (mm)
Weight73.5g, e nang le betri
motlakaseng3V DC, 50mA, 2 AAA alkaline batteries
memoriLiphetho tsa liteko tsa tsoekere ea mali tse 10
Boemo ba ts'ebetso10 ℃ ~ 35; ≤80% RH
kahoE tšoeroe ka letsoho
Mealo ea litekanyomg / dL kapa mmol / L
Mealo ea tekanyo20 ~ 600 mg / dL kapa 1.1 ~ 33.3mmol / L
Hlakola bopheloLilemo tse 10 (e hakantsoe ke liteko makhetlo a 7 ka letsatsi).
Nakong ea ts'ebeliso, mosebelisi o lokela ho boloka sehlahisoa
lebisa litlhoko tsa buka ena ea mosebelisi.
Evalu Tlhahlobo ea ts'ebetso ea mosebelisi:

Results for glucose concentrations<100 mg/dL (<5.55mmol/L)
Ka har'a ± 5mg / dL
(within ± 0.28mmol/L)
Ka har'a ± 10mg / dL
(within ± 0.56mmol/L)
Ka har'a ± 15mg / dL
(within ± 0.83mmol/L)
56% (5/9)100% (9/9)100% (9/9)
Results for glucose concentrations ≥ 100 mg/dL (≥5.55mmol/L)
Ka har'a ± 5%
71% (65/91)
Ka har'a ± 10%
93% (85/91)
Ka har'a ± 15%
98% (89/91)