Green-Yellow-Red light warning;

Auto sample loading detection;

Auto electrode inserting detection;

Affordable;Code free; CE certification

шарҳ
  • Истифодаи истифодашуда

    Safe Accu2 is intended for IVD testing the blood glucose in the fresh whole capillary blood and vein plasma; for medical institutes’ fast testing of blood glucose; for diabetics and other groups’ blood glucose monitoring.

  • Принсипи санҷиш

    C6H12O6 + K3Fe(CN)6   →   FAD-GDH   →  C6H12O7 + K4Fe(CN)6
                        K4Fe(CN)6 → K3Fe(CN)6+ e-

  • Хусусиятҳо (Афзалиятҳои техникӣ)

    1) FAD-GDH system : It is not easily affected by oxygen molecule in the blood.

    2) User-friendly operation system: no coding & automatic test strip ejectionin the management of diabetes.

Мушаххасоти
Ҳаҷми хун 0.6μL
Намуди намуна Тамоми хуни хунрези капилляр
Calibration Муодили плазма
Вақтсанҷ 10
Ҳолати нигоҳдорӣ / интиқолдиҳанда -20C ~ 55 ℃
андоза 105 * 60 * 23 (мм)
тарозу 77.2g,with battery
манбаи Ҳокимият 3V DC,10mA,2 AAA alkaline batteries
хотира 200 натиҷаи ташхиси глюкозаи хун
Ҳолати корӣ 10 ℃ ~ 35C≤80% RH
Сохтмон Дастӣ
Воҳидҳои ченкунӣ мг / дл ё ммол / л
Миқдори андозагирӣ 20~600 mg/dL or 1.1~33.3mmo/L
Precision: test result of Safe-Accu2 meet the requirement below:

Accuracy: 95% test result of Safe-Accu 2 meet the requirement below:


Диапазони консентратсия Хатои%
<5.5 ммоль / л (100 мг / дл) Within±0.83 mmol/L(15 mg/dL)
25.5 ммоль / л (100мг / дл) Дар доираи ± 15%


Precision: test result of Safe-Accu 2 meet the requirement below:


Диапазони консентратсия Талаб мент
<5.5 ммоль / л (100 мг / дл) SD<0.34mmol/L(6.0mg/dL)
25.5 ммоль / л (100мг / дл) CV <6.0%



                         