I-Accu2 ekhuselekileyo
Green-Yellow-Red light warning;
Auto sample loading detection;
Auto electrode inserting detection;
Affordable;Code free; CE certification

isishwankathelo
Se tyenziswa
Safe Accu2 is intended for IVD testing the blood glucose in the fresh whole capillary blood and vein plasma; for medical institutes’ fast testing of blood glucose; for diabetics and other groups’ blood glucose monitoring.
Vavanya uMgaqo
C6H12O6 + K3Fe(CN)6 → FAD-GDH → C6H12O7 + K4Fe(CN)6
K4Fe(CN)6 → K3Fe(CN)6+ e-Iimpawu (iinzuzo zobugcisa)
1) FAD-GDH system : It is not easily affected by oxygen molecule in the blood.
2) User-friendly operation system: no coding & automatic test strip ejectionin the management of diabetes.
Specification
Umthamo wegazi | 0.6μL |
Uhlobo lwesampulu | I-Capillary igcwele igazi elipheleleyo |
Ukulinganisa | Iplasma ilingana |
Ixesha lokulinganisa | 10s |
Ukugcinwa kweemitha / imeko yokuthutha | -20C ~ 55℃ |
Dimension | I-105 * 60 * 23 (mm) |
ubunzima | 77.2g, enebhetri |
Power umthombo | 3V DC,10mA,2 AAA alkaline batteries |
Imemori | Iziphumo zovavanyo lweswekile yegazi ezingama-200 |
Imeko yokusebenza | 10℃~35C≤80% RH |
ukwakhiwa | Ebanjwe ngesandla |
Iiyunithi zokulinganisa | mg / dL okanye mmol / L |
Uluhlu lwesilinganiselo | 20–600 mg/dL okanye 1.1–33.3mmo/L |
Precision: test result of Safe-Accu2 meet the requirement below:
Accuracy: 95% test result of Safe-Accu 2 meet the requirement below:
Uluhlu loxinzelelo | Ukwahlula% |
---|---|
ngaphantsi kwe-5.5 mmol/L(100 mg/dL) | Within±0.83 mmol/L(15 mg/dL) |
25.5 mmol/L(100mg/dL) | Ngaphakathi ± 15% |
Precision: test result of Safe-Accu 2 meet the requirement below:
Uluhlu loxinzelelo | Funa ingcebiso |
---|---|
ngaphantsi kwe-5.5 mmol/L(100 mg/dL) | SD<0.34mmol/L(6.0mg/dL) |
25.5 mmol/L(100mg/dL) | I-CV <6.0% |