I-Accu2 ekhuselekileyo

Green-Yellow-Red light warning;

Auto sample loading detection;

Auto electrode inserting detection;

Affordable;Code free; CE certification

isishwankathelo
  • Se tyenziswa

    Safe Accu2 is intended for IVD testing the blood glucose in the fresh whole capillary blood and vein plasma; for medical institutes’ fast testing of blood glucose; for diabetics and other groups’ blood glucose monitoring.

  • Vavanya uMgaqo

    C6H12O6 + K3Fe(CN)6   →   FAD-GDH   →  C6H12O7 + K4Fe(CN)6
                        K4Fe(CN)6 → K3Fe(CN)6+ e-

  • Iimpawu (iinzuzo zobugcisa)

    1) FAD-GDH system : It is not easily affected by oxygen molecule in the blood.

    2) User-friendly operation system: no coding & automatic test strip ejectionin the management of diabetes.

Specification
Umthamo wegazi0.6μL
Uhlobo lwesampuluI-Capillary igcwele igazi elipheleleyo
UkulinganisaIplasma ilingana
Ixesha lokulinganisa10s
Ukugcinwa kweemitha / imeko yokuthutha-20C ~ 55℃
DimensionI-105 * 60 * 23 (mm)
ubunzima77.2g, enebhetri
Power umthombo3V DC,10mA,2 AAA alkaline batteries
ImemoriIziphumo zovavanyo lweswekile yegazi ezingama-200
Imeko yokusebenza10℃~35C≤80% RH
ukwakhiwaEbanjwe ngesandla
Iiyunithi zokulinganisamg / dL okanye mmol / L
Uluhlu lwesilinganiselo20–600 mg/dL okanye 1.1–33.3mmo/L
Precision: test result of Safe-Accu2 meet the requirement below:

Accuracy: 95% test result of Safe-Accu 2 meet the requirement below:


Uluhlu loxinzeleloUkwahlula%
ngaphantsi kwe-5.5 mmol/L(100 mg/dL)Within±0.83 mmol/L(15 mg/dL)
25.5 mmol/L(100mg/dL)Ngaphakathi ± 15%


Precision: test result of Safe-Accu 2 meet the requirement below:


Uluhlu loxinzeleloFuna ingcebiso
ngaphantsi kwe-5.5 mmol/L(100 mg/dL)SD<0.34mmol/L(6.0mg/dL)
25.5 mmol/L(100mg/dL)I-CV <6.0%



                         