Agbọn Smart AQ ailewu
Fast 5s test time; FDA-GDH monitoring system; SINODRAW patent lancing device; Temperature warning

Akopọ
Akopọ
The Safe AQ Smart Blood Glucose Monitoring System is designed for the quantitative measurement of glucose in fresh capillary whole blood samples taken from the fingertip and in venous whole blood samples. The Blood Glucose Monitoring System is for use outside the body only (in vitro diagnostic use) for self-testing and professional use as an aid in the management of diabetes.
Ofin idanwo
A blood glucose test is based on measurement of electrical current caused by the reaction of glucose with the reagents (special chemicals) on the electrode of the strip. The blood or control solution sample is drawn into the tip of the test strip through capillary action. Glucose in the sample reacts with the special chemicals and generates electrons, which produce electrical current. The blood glucose meter measures the electrical current and calculates the glucose result. The glucose result is displayed by your meter as mg/dL or mmol/L.
Awọn ẹya (Awọn anfani Imọ-ẹrọ)
1) Eto FAD-GDH: Ko ni rọọrun nipasẹ molikula atẹgun ninu ẹjẹ.
2) Eto ṣiṣe ti olumulo-ore: ko si ifaminsi & ifa idanwo ifilọlẹ adactionin iṣakoso ti àtọgbẹ.
Specification
Iwọn ẹjẹ | 0.6μL |
Ayẹwo apẹẹrẹ | Fi ẹjẹ gba gbogbo ẹjẹ |
odiwọn | Pilasima deede |
Akoko wiwọn | 5s |
Ibi ipamọ Mita / ipo gbigbe | -20℃~ 55 ℃ |
apa miran | 84 * 60 * 26 (mm) |
àdánù | 73.5g, with battery |
orisun agbara | 3V DC, 50mA, awọn batiri alkaline 2 AAA |
Memory | Awọn abajade idanwo glukosi ẹjẹ |
Ipo iṣiṣẹ | 10 ℃ ~ 35℃; ≤80% RH |
ikole | Ọwọ-waye |
Awọn iwọn wiwọn | mg / dL tabi mmol / L |
Iwọn wiwọn | 20 ~ 600 mg / dL tabi 1.1 ~ 33.3mmo / L |
selifu aye | Awọn ọdun 10 (iṣiro nipasẹ idanwo awọn akoko 7 fun ọjọ kan). Lakoko lilo, olumulo yẹ ki o ṣetọju ọja naa tọka si awọn ibeere olumulo yii. |
Awọn iṣẹ iṣe iṣe
Accuracy: 95% test result of Safe AQ meet the requirement below:
Ibiti Ikanju | Bias% |
---|---|
<5.5 mmol / L (100 mg / dL) | Laarin ± 0.83 mmol / L (15 mg / dL) |
25.5 mmol / L (100mg / dL) | Laarin ± 15% |
Precision: test result of Safe AQ Smart meet the requirement below:
Ibiti Ikanju | Beere ment |
---|---|
<5.5 mmol / L (100 mg / dL) | SD <0.34mmol / L (6.0mg / dL) |
25.5 mmol / L (100mg / dL) | CV <6.0% |