Bi-functional blood glucose and uric acid monitoring system;  Automatic strips ejection; ISO15197 2013 approved;  Smart code for uric acid test

Akopọ

      The Safe AQ UG Blood Glucose and Uric Acid Monitoring System is designed for the quantitative measurement of glucose and uric acid in fresh capillary whole blood samples and in venous whole blood samples. The Safe AQ UG Blood Glucose and Uric Acid Monitoring System is for use outside the body only (in vitro diagnostic use) for self-testing and professional use as an aid in the management of diabetes and hyperuricemia (HUA).

Specification
Iṣe IdanwoẸjẹ GlucoseUric Acid
igbeyewo ọnaFAD-GDHOxidized Ferrocene itọsẹ
Iru apẹẹrẹẸsẹ keekeeke odidi ẹjẹ / ẹjẹ gbogbo iṣanẸsẹ keekeeke odidi ẹjẹ / ẹjẹ gbogbo iṣan
Ẹjẹ ẹjẹ
0.6µl3µl
Ipele idanwo20-600mg / dL
 (1.1-33.3mmol / L)
3.0-20.0mg / dL
(181-1188 µmol/L)
Akoko Idanwo5s25
Memory500 tosaaju100 tosaaju
Wiwulo rinhoho24 osu12 osu
Evalu Iṣiro iṣẹ olumulo glukosi:
Awọn abajade fun awọn ifọkansi glucose <100 mg / dL (<5.55mmol / l)
Laarin ± 5mg / dL
(laarin ± 0.28mmol / l)
Laarin ± 10mg / dL
(laarin ± 0.56mmol / l)
Laarin ± 15mg / dL
(laarin ± 0.83mmol / l)
50% (14/28)85.71% (24 / 28)100% (28 / 28)
                Awọn abajade fun awọn ifọkansi glucose ≥ 100 mg / dL (≥5.55mmol / l)
Laarin ± 5%
Laarin ± 10%
Laarin ± 15%
55.56% (40/72)90.28% (65/72)97.22% (70/72)
◆ Uric acid user performance evaluation:
Results for uric acid concentrations < 297μmol/L (5.0 mg/ dL)
Within ± 14.9μmol/L
(Within ± 0.3 mg/ dL)
Within ± 29.7μmol/L
(Within ± 0.5 mg/ dL)
Within ± 59.4μmol/L
(Within ± 1.0 mg/ dL)
40.0% (6/15)60.0% (9 / 15)100% (15 / 15)
                Results for uric acid concentrations≥ 297μmol/L (5.0 mg/ dL)
Laarin ± 5%
57.6% (49/85)
Laarin ± 10%
94.1% (80/85)
Laarin ± 15%
100% (85/85)





Pe wa