I-Accu2 ephephile
Green-Yellow-Red light warning; Auto electrode inserting detection; Auto sample loading detection; Affordable;Code free; CE certification

Uhlolojikelele
Ukusetshenziswa okuhloswe
Safe Accu2 is intended for IVD testing the blood glucose in the fresh whole capillary blood and vein plasma; for medical institutes’ fast testing of blood glucose; for diabetics and other groups’ blood glucose monitoring.
Isimiso Sokuhlola
C6H12O6 + K3Fe(CN)6 → FAD-GDH → C6H12O7 + K4Fe(CN)6
K4Fe(CN)6 → K3Fe(CN)6+ e-Izici (Izinzuzo zobuchwepheshe)
I-1) Uhlelo lwe-FAD-GDH: Aluthinteki kalula i-molecule ye-oxygen esegazini.
I-2) Uhlelo lokusebenza lokusebenziseka kalula: akukho khodi yokufaka amakhodi nokufaka i-othomathikhi yokuhlolwa kwesifo sikashukela.
Ukucaciswa
Umthamo wegazi | I-0.6μL |
Uhlobo lwesampula | Igazi eliphelele eligcwele igazi |
Ukulinganisa | I-Plasma ilingana |
Isikhathi sokulinganisa | 10s |
Isimo sokugcina imitha / yokuhamba | -20C ~ 55 ℃ |
Dimension | I-105 * 60 * 23 (mm) |
Isisindo | 77.2g, nebhethri |
umthombo Power | 3V DC,10mA,2 AAA alkaline batteries |
Memory | Imiphumela yokuhlolwa kweglucose ye-200 |
Isimo sokusebenza | I-10 ℃ ~ 35C≤80% RH |
Ukwakhiwa | Ibanjwe ngesandla |
Amayunithi wokulinganisa | mg / dL noma mmol / L |
Uhla lokulinganisa | I-20 ~ 600 mg / dL noma i-1.1 ~ 33.3mmo / L |
Precision: test result of Safe-Accu2 meet the requirement below:
Accuracy: 95% test result of Safe-Accu 2 meet the requirement below:
Ububanzi Bokubhekisisa | I-Bias% |
---|---|
<5.5 mmol / L (100 mg / dL) | Ngaphakathi ± 0.83 mmol / L (15 mg / dL) |
I-25.5 mmol / L (100mg / dL) | Ngaphakathi ± 15% |
Precision: test result of Safe-Accu 2 meet the requirement below:
Ububanzi Bokubhekisisa | Dinga i-ment |
---|---|
<5.5 mmol / L (100 mg / dL) | I-SD <0.34mmol / L (6.0mg / dL) |
I-25.5 mmol / L (100mg / dL) | I-CV <6.0% |